Simulated use test, Biotech-germande Laboratory on Soluscope 3 GTA:
Results on Standard Cycle n°1 (20 minutes, including 5 min disinfection at 45°C)

     
Strains

Mean log reductions, after a manual pre-treatment

Min value
Average
Max value
 
With a Pretreatment
Pseudomonas aeruginosa
8.8
9.3 ± 0.6
10.0
Aspergillus niger
5.6
6.8 ± 0.7
7.5
Bacillus subtilis
5.9
6.5 ± 0.4
7.1
Bacillus cereus
5.7
6.2 ± 0.3
6.6
Mycobacterium terrae
7.0
7.5 ± 0.3
8.0
 
Strains

Nb successfull tests / Nb performed tests, after a manual pre-treatment, to reach a disinfection level

standard
reinforced
Pseudomonas aeruginosa
6/6

SCORE (*)

 

5 / 5

6/6

SCORE (*)

 

4.5 / 5

Aspergillus niger
5/6
5/6
Bacillus subtilis
6/6
5/6
Bacillus cereus
6/6
5/6
Mycobacterium terrae
6/6
6/6

(*) : reached score = (nb of successful tests / nb of performed tests) x 5

For information:
Strains

Minimal Reductions required to reach a :

Standard Disinfection Level
Reinforced Disinfection Level
Pseudomonas aeruginosa
Aspergillus niger
Bacillus subtilis
Bacillus cereus
Mycobacterium terrae
7 log
6 log
5 log
4 log
4 log
7 log
6 log
5 log
6 log