Soluscope 2
|
Microbiological Efficacy
|
| |
|
|
| Soluscope D In vitro tests |
|
| |
Soluscope chemistry and process have been tested by the most famous laboratories according to
the French, British, European, Australian and US standards.
The minimum disinfectant concentration used in the Soluscope is 0,2% whilst the Minimum effective
Concentration (MEC) determined by the tests is 0,15%.
Most of the tests were done in worse case conditions: product at the end of its shelf life,
below the minimum used concentration, with organic soil and hard water, in order to reinforce the margin of safety.
Summary of in vitro test
|
|
|
| Soluscope process in use tests |
|
| |
The in use tests evaluate the process efficacy by using artificially contaminated endoscopes that are sampled after being
reprocessed in the Soluscope.
The methods used are slightly different (inoculum, type of sampling liquid…) but they all placed Soluscope
in the worst possible conditions: contamination of all channels with an inoculum > 106 microorganisms per ml with
interfering substances, without precleaning of the endoscope, waiting until the contamination
material has dried into the channels.
Soluscope has successfully passed the French Germande evaluation and the British one from HIRL
|
|
| |
|
|
| Self disinfection evaluation test |
|
| |
|
Self
disinfection test done by the Germande One
of the most encountered troubles with washer disinfectors
is their big risk to get contaminated and to be a potential
source of endoscope infection. Soluscope has been the
first washer disinfector to have its self-disinfection
capability tested. A self-contamination occurs for 3
main reasons : a failure or a back flow contamination
of the water treatment system (filter, UV, etc) that
contaminates the machine, a misconception of the hydraulic
circuit (dead ends) that prevents a total disinfection
of the machine, or a contamination of the machine from
the endoscope.
|
|
|
|
The complete scientific file of Soluscope